GF1061PT with Polarity, ratio & phase error Portable Voltage transformer PT Analyzer
GF1061PT is the first portable device (4.8 kg) which offers highly accurate voltage transformer tests. This allows to use GF1061PT not only for electrical performance checks, but also for class verification and calibration. It performs quick tests of all kinds of inductive voltage transformers (VTs or PTs) and capacitive voltage transformers (CVTs) for both protection and metering purposes. Its lightweight design makes it ideal for on-site tests and calibration tasks in power system grids.
As a manufacturer or testing lab you can use GF1061PT in your production facilities and test/development labs. it can finish the measurements (M) and protection (P) class PT. Adopt 7 inch touch TFT color LCD, self-equipped mini type printer supporting field printing; supporting to use USB flash disk to dump data. The GF1061PT portable PT tester is the most complete and easy-to-use testing system for PTs according to IEEE and IEC standards.
I. Voltage Transformer (PT) |
---|
1. Excitation characteristic test |
2. Transformation ratio test |
3. Polarity |
4. Ratio error, phase error |
5. Degauss |
6. Calculation of knee point value |
7. Actual secondary load test (burden test) |
8. Resistance test |
Electrical parameters | ||
---|---|---|
Accuracy | 0.02% | |
Power supply | AC 220V±10% or AC 120V±10%, 50/60Hz or Battery | |
Output voltage | 0-100Vrms | |
Output current | 0-5Arms (20A peak-value) | |
Output power | 0-400 VA (1500 VApeak) | |
Automatic frequency variation range | 0.1-60Hz | |
Equivalent excitation voltage | ≤5000V/50KV | |
Accuracy | ≤0.5% | |
Secondary winding DC resistance measurement | Range | 0.1-1000Ω |
Accuracy | ≤0.05% | |
Secondary actual load measurement | Range | 5VA-1000VA |
Accuracy | ≤0.5%±0.1VA | |
PT phase error measurement | Accuracy | ±1min (typical) / 3 min (guaranteed) |
Resolution | 0.1min | |
PT ratio error measurement | Range | 1-15000 |
Accuracy | ≤0.05% | |
LCD display | 7' inch TFT touch color LCD | |
Communication port | USB port | |
Printer | Yes, Thermal printer | |
Standards | ||
Reference standards | GB1207-2006, GB1208-2006, GB16847-1997 IEC60044-1, IEC60044-6, IEC61869, ANSI/IEEE C57.13 | |
Safety standards | GB 4793.1-2007 | |
EMC | EMC standard 89/336/EEC | |
FCC Subpart B of Part 15 Class A | ||
IEC 1000-4-2/3/4/6 | ||
Mechanical parameters | ||
Overall dimension (L x W x H) (mm) | 280 x 250 x 160 | |
Weight (kg) | ≤4.8 | |
Environmental conditions | ||
Relative humidity | Relative humidity 5%-95% not condensing | |
Operating temperature | -10°C to +50°C | |
Storage temperature | -20°C to +70°C | |
Altitude | ≤1000m |
Beijing GFUVE Instrument Transformer Manufacturer Co.,Ltd.
Beijing GFUVE Instrument Transformer Manufacturer Co., Ltd. was established in 2008. It combined design, manufacturing and sales together as one high-tech enterprise. GFUVE Instrument Transformer company mainly produces less than 35kV current transformer and voltage transformer, split core ct and current clamp models etc.
GFUVE is a leader manufacturer of split core current transformer products in China. It is professional in current transformer, potential transformer and related accessory. The product of outdoor split core current transformer develops very quickly in the global market. Our company has achieved certification to ISO9001. The products are sold to over 30 provinces in China and exported to overseas including South America, Middle East, Southeast Asia, Australia and Africa.
Please contact us now for further business relationship. We are here to provide you the best service and the best product.